English | French | Espanol | Italiano | Portugues | Deutsch | Arabic

ICU Equipments

Pages: 1 2 [3] [«Back ]
product/ICU Equipments/pg74_2.jpg

CARDIART 8408 VIEW

Product Code: AVI-1556
Features:
  1. Automatic, manual and pre-programmed recording mode
  2. Multi-channel print format on A4 paper : 3, 6x1, 6x2, “Full Page”(3x4+R) and (3x4+3R), 12 channel
  3. LCD display for ECG real time 3, 6, 6+6 channels
  4. Full Alphanumeric keyboard
  5. Rhythm lead recording
  6. Qualitative and quantative analysis
  7. Optional software for PC storage or real time ECG monitoring
  8. Built-in rechargeable battery

 

 
product/ICU Equipments/pg74_4.jpg

Defibrillator

Product Code: AVI-1557
Features:
  1. 5S to charge to 360J with battery
  2. 100 charge/discharge with fully charged new battery
  3. Recording of past 28 events
  4. Thermal printing
  5. Slide on Adult paddle over paediatric paddles
  6. LCD for sharp bright non fade display

 

 
product/ICU Equipments/pg74_5.jpg

Defibrillator

Product Code: AVI-1558
Features:
  1. Light weight, portable design (weighs less than 2Kg)
  2. Easy to use, simple 1,2,3 operation
  3. Large, easy-to-read screen
  4. Innovative SCOPE escalating biphasic waveform for high defibrillation efficancy
  5. Optional rechargeable battery operations
  6. Suitable for para medics

 

 
product/ICU Equipments/pg74_3.jpg

HALTER RECORDER

Product Code: AVI-1559
Features:
  1. Compact, lightweight & handy
  2. Accurate & reliable recordings
  3. 3 channel ECG recording
  4. ‘giOtto’ holter analysis & ‘Holter Pilot’ exam management software
  5. Seperate channel for pace maker pulses

 

 
product/ICU Equipments/1559.jpg

ANAESTHESIA VENTILATOR – MX-10AL

Product Code: AVI-1560
  • Easy to operate
  • Compatible with anaesthesia machine
  • Standalone-no need of driving gas
  • Portable
  • Suitable for use in ambulance
  • Ultra low operating cost
  • Low maintenance
  • Excellent for use in operation theater
  • Thousands of satisfied customers
  • Low cost of ownership

  SPECIFICATIONS

Operational Parameters

 

 

 

Operating Principle

Time Cycled,Volume Adjustable

Models

CMV

Breaths per minute

6-60

Tidal Volume

50-1100 cc

Max.inspiratory pressure

60 cm H2O

Flow Pattern

Constant Flow

I:E Ratio

4:1 – 1:4

Patient Safety

Adjustable pressure Release valve

 

 

Electrical Supply

 

 

 

Input Supply

184-264 V A.C 50/60 Hz

Power Consumption

50 VA

 

 

Physical Specifications

 

 

 

Dimensions

485 [L] x 290 [W] x 250 [H] mm

Weight

13 kgs.

 

 

Alarms [Audio/Visual]

Ventilot Malfunction

 

Patient Disconnection

 

Excessive Inlet Gas Flow

 

Insufficient Inlet Gas Flow

 

High Inspiratory Pressure [adjustable 0-50 cm H2O

 

Low Inlet Volume

 

Power Failure

Accessories

Basic Breathing Circuit Set [Single Use]


 

 
product/VENTILATORS/4.jpg

MX-30 Critical CareVentillator

Product Code: AVI-1560 A
offers microprocessor based technology in a compact,light weight,portable ventillator syystem. Ventillators offer ease of operation,high functional reliability and consisitency.They are totally electrically driven and do not require compressed gas or air.This feature makes ventillators highly suitable for portable,mobile and other applications where centralized gas is not available.Various monitoring alarms enable their usage in unattended applications.
   
MX-30-Technical specifications   Anaesthesia Ventillator
Features
  • Easy to operate,minimum controls.
  • Light Weight.
  • Excellent for use in operation theatre.
  • Electrically operated,does not require compressor or gas,
  • Low maintenance.
  • Low operating cost.
  • Compatible with anaesthesia amchine.
  • Portable-suitable for use in an ambulance.
Specifications :
Operating Principle Microprocessor Controlled and Electrically driven ventilator.
Input Supply 184-264 V,50/60Hz,A.C.
Power Consumption  50VA.
Weight 13Kgs.
Dimension(LXWXH)  (485*290*250)mm.
Operational Parameters:
Modes CMV
Breathes Per Minute 6-60
Tidal Volume 50-100cc
Maximum Inspiratory Pressure 60cm H20
Flow Pattern Linear
I:E Ratio 4:1 – 1:4
   
Alarms(Audio-visual) Accessories
Ventillator Malfunction Basic Breathing Circuit Set  ( single use) 
Patient Disconnection. Excessive Inlet Gas Flow Patient Valve-Fish Mouth Type (Reusable)
Insufficient Inlet Gas Flow Power Cord
High Inspiratory Pressure Spare Fuses (adjustable from 0-50cm H20)
Low Inlet Volume  
Power Failure  

 

 
product/VENTILATORS/mx40p.jpg

MX-40p Critical Care Ventilator

Product Code: AVI-1561

MAX MEDITECH offers microprocessor based technology in a compact, light weight, portable ventilator system. Max Ventilators offer ease of operation, high functional reliability and consistency. They are totally electrically driven and do not require compressed gas or air. This feature makes Max ventilators highly suitable for portable, mobile and other applications where centralized gas is not available. Various monitoring alarms enable their usage in unattended applications.

  • Easy to operate, minimum controls
  • Excellent for use in operation theatre
  • Electrically operated, does not require compressor or gas
  • Low maintenance
  • Low operating cost
  • Portable - suitable for use in an ambulance
  • Application: Anaesthesia ICCU and Ambulance
Specifications
Operating Principle Microprocessor controlled and electrically driven ventilator
Input Supply 184-264 V, 50/60 Hz, A.C
Power Consumption 50 VA
Weight 14 Kgs
Dimension (LXWXH) (485 X 290 X 250) mm
Operational Parameters
Modes CMV, ASSIST, SIMV
Breaths per Minute 1 - 60 (CMV : 6-60, ASSIST/SIMV : 1-60)
Tidal Volume 50 - 1100 cc
Maximum Inspiratory Pressure 60 cm H2O
Flow Pattern Linear
I:E Ratio 4:1 - 1:4
Inspiratory Time 0.06 - 4 secs
Pressure Trigger -9 to -1 cm H2O w.r.t atmosphere
Alarms
  • Patient Disconnection
  • Excessive Inlet Gas Flow
  • Insufficient Inlet Gas Flow
  • High Inspiratory Pressure (Adjustable from 0-50 cm H2O)
  • Apnea
  • Low Inlet Volume
  • Ventilator Malfunction
  • Power Failure

 

Accessories
  • Basic Breathing Circuit Set (Single Use)
  • Patient Valve - Fish Mouth Type (Reusable)
  • Power Cord
  • Trolley
  • Spare Fuses

 

 
product/VENTILATORS/protonventilator.jpg

Proton Intensive Care Ventilator

Product Code: AVI-1562

PROTON is a full spectrum intensive care ventilator for Paediatric and Adult patients and provides complete ventilation solutions. PROTON is pneumatically powered-micro Processor Controlled ventilator. It has all advance invasive and non invasive (Bi-level) modes with facility of showing Pressure, Volume and Flow graphs. O2 blending and built in nebulizer are another distinct advantages of PROTON. It comes with colour TFT display.

 
Operartional Parameters
Modes CMV - Volume Controlled
  SIMV - Volume Controlled
  ASSIST - Apnea Back-Up
  PSV-SIMV
 

PSV - Apnea Back-Up

  CMV - PCV / ASSIST - PCV / SIMV - PCV
  Bi - NIV / CPAP
 
Displayed Parameters

Breaths per Minute (BPM)

0 to 120

Tidal Volume (VT)

30 to 300 cc in Pediatric Mode

300 to 3000 cc in Adult Mode

Inspiratory (VT)

0.16 to 4 seconds or 20 to 80% of Breath cycle

Peak Flow Rate Pediatric:
Adult:
6 to 50 lit./min maximum
6 to 150 lit./min maximum
Inspiratory Pressure 0 to 80 cm of H20
PEEP 0 to 35 cm of H20
Inspiratory Pause 15 to 50% of Ti
O2 Blending 21 to 100%
Pressure Trigger .5 to 15 cm of H20 below PEEP
Flow Trigger 0.3 to 30 Ipm
Apnea Back-Up Time 5 to 60 sees
 
Alarms : Preset and Settable
Airway Pressure High, Low
Inspired VT High, Low
Expired Volume Low
FiO2 High, Low
PEEP High, Low
Breaths High
Apnea Alarm Time 5 to 60 seconds
Mechanical Pressure Release Valve 90 cm of H2O
 
Measured Parameters
No of Breaths Spontaneous, Total Breaths
per minute
Airway Pressure Peak, Plateau, Mean, PEEP
Patient Trigger Effort Trigger Pressure, Trigger Flow
Tidal Volume Inspired, Expired
O2 Concentration Inspired O2
Graphics on 57" LCD Pressure-Time / Volume-Time / Flow-Time
 
Electrical and Gas Supply
Input Supply 175 to 260 Volts OR 90 to 132 Volts AC, 50/60 Hz
Internal Battery Back-Up 60 Minutes
External Battery Optional : 10 to 15 Volts DC
Power Consumption 70 Watts (Approx.) for basic ventilator
VA Requirement I I5 VA (Approx.) for basic ventilator
Gas Input Medical Grade Air & O2 at 2.7 to 6 bar (39 to 87 psi)
 
Physical Specifications
Dimensions 480 (W) x 290 (D) x 285 (H) mm without trolley
Weight 15 Kg for basic ventilator unit
Display 5.7" Colour TFT LCD (Internal)
Communication Output RS-485 connectivity for monitoring multiple units
 
Accessories  
Breathing Circuit (Single Use) External TFT Monitor
Nebuliser Unit Tube Support Arm
Fluid Collection Bottl Compressor (Optional)
HME, Bacteria Filter Humidifier (Optional)
Trolley (Cart) NIV-Mask (Optional)
 
External TFT LCD for Enhanced Monitoring
Parameters Dynamic Airway Pressure Indication
Mode, BPM, Spont. BPM,Vti ,Vte, PEEP, Ppeak, FiO2
Graphical Display of Lung Mechanics Airway Pressure v/s Time
Tidal Volume v/s Time
Flow Rate v/s Time
Airway Pressure v/s Flow Rate
Tidal Volume v/s Airway Pressure
Flow Rate v/s Tidal Volume
Display Size 47 cm Wide Screen TFT LCD Monitor
Resolution 1024 x 768 Pixels
Power Consumption 30 Watts approximately
Supply Voltage 230 Volts A.C., 50 Hz

 

 
product/VENTILATORS/zeon.jpg

Zeon Critical Care Ventilator

Product Code: AVI-1563
MAX MEDITECH offers microprocessor based technology in a compact, self contained ventilator system. Max ventilators offer ease of operation, high functional reliability and consistency. They are totally electrically driven and do not require compressed gas or air. This feature proves Max ventilators highly suitable for hospitals and nursing homes where centralized gas supply is not available.
  • Suitable for critical care applications
  • Microprocessor controlled, electrically driven ventilator
  • Large LCD display
  • In-built battery backup
  • User friendly controls
  • Low operating cost
  • Application: ICU and ICCU
Specifications
Input Supply 185 - 264 V, 50/60 Hz, AC
Power Consumption 40 Watts
VA Requirement 70 VA
Internal Battery 2 X 6 V, 4.5 AH
Battery Back-up Duration 60 minutes
Dimensions 480 mm (W) x 350 mm (D) x 285 mm (H)
Weight 20 Kg
Operating Temperature 10 - 40oC
 
Operational Parameters
Modes CMV-VC, ASSIST-VC, SIMV-VC, PSV-ASSIST, SIMV-PSV,PSV APB, CMV-PCV, ASSIST-PCV, SIMV-PCV
Breaths per Minute (1 to 60 - SIMV) 6 to 60
Tidal Volume 30 - 1100 cc
Inspiratory Pressure 10 to 60 cm H2O
Flow Pattern Constant
I:E Ratio 1:4 to 4:1
Inspiratory Time 0.04 to 4 sec (20% to 80% in CMV)
Pressure Trigger -15 to -0.5 cm H2O
Flow Trigger 1.8 to 30 lpm
Peep Level 0 to +20 cm H2O
PCV Limit 10 to 60 cm H2O
PSV LImit 10 to 60 cm H2O
Apnea Backup Time 5 to 60 secs
Other Features Inspiratory Pause Limit, Sigh Breath, Manual Breath
 
Alarms/Indications
  • Excess Breaths
  • Low Expired Volume
  • Low Inspired Volume
  • FiO2 Sensor Open
  • FiO2 Sensor Defective
  • FiO2 High
  • FiO2 Low
  • High Inspiratory Pressure
  • Low Inspiratory Pressure
  • Patient Disconnection
  • Ventilator Malfunction
  • Low Inlet Volume
  • Check Exp Flow Sensor
  • Apnea
  • Power Failure
  • Peep Low
  • Peep High
  • Battery Operation
  • Battery Low/ Battery Discharged
  • Ventilator Failure
   
Monitored Parameters  
  • FiO2 Monitoring
  • Spontaneous Breaths
  • Inspired Tidal Volume
  • Expired Tidal Volume
  • Peak Inspiratory Pressure
  • Plateau Pressure
  • PEEP Level
  • Trigger Level
  • Delivered BPM
   
Connections  
  • 22 mm (Male) outlet port for tubing connection to patient.
  • 22 mm (Male) expiratory port for tubing connection from patient.
  • 8 mm low pressure connection for oxygen enrichment.
  • Pressure monitoring port.
  • Reservoir Bag Port
 

 

 
Pages: 1 2 [3] [«Back ]
 
   
 
   

Operation Theatre Equipments

Hospital Furnitures

Hospital Hollowares

Pathology Instruments

Plastic Wares

Blood Bank Instruments

ICU Equipments

Diagnostic Instruments

Anaesthesia & Critical Care Ventilators

Rubber Products

OPD Instruments

General Instruments

ENT OPD Unit

Surgical Instruments

Orthopaedic Implants

Physiotherapy Equipments

Rehabilitation Aid

X-Ray Equipments

Paediatric Instruments

Psychology Apparatus

Opthalmic Instruments

Surgical Disposables

Dental Instruments

Mortuary Chamber

Projectors

Physiological Models