| MAX MEDITECH offers microprocessor based technology in a compact, self contained ventilator system. Max ventilators offer ease of operation, high functional reliability and consistency. They are totally electrically driven and do not require compressed gas or air. This feature proves Max ventilators highly suitable for hospitals and nursing homes where centralized gas supply is not available. |
- Suitable for critical care applications
- Microprocessor controlled, electrically driven ventilator
- Large LCD display
- In-built battery backup
- User friendly controls
- Low operating cost
- Application: ICU and ICCU
|
| Specifications |
| Input Supply |
185 - 264 V, 50/60 Hz, AC |
| Power Consumption |
40 Watts |
| VA Requirement |
70 VA |
| Internal Battery |
2 X 6 V, 4.5 AH |
| Battery Back-up Duration |
60 minutes |
| Dimensions |
480 mm (W) x 350 mm (D) x 285 mm (H) |
| Weight |
20 Kg |
| Operating Temperature |
10 - 40oC |
| |
| Operational Parameters |
| Modes |
CMV-VC, ASSIST-VC, SIMV-VC, PSV-ASSIST, SIMV-PSV,PSV APB, CMV-PCV, ASSIST-PCV, SIMV-PCV |
| Breaths per Minute |
(1 to 60 - SIMV) 6 to 60 |
| Tidal Volume |
30 - 1100 cc |
| Inspiratory Pressure |
10 to 60 cm H2O |
| Flow Pattern |
Constant |
| I:E Ratio |
1:4 to 4:1 |
| Inspiratory Time |
0.04 to 4 sec (20% to 80% in CMV) |
| Pressure Trigger |
-15 to -0.5 cm H2O |
| Flow Trigger |
1.8 to 30 lpm |
| Peep Level |
0 to +20 cm H2O |
| PCV Limit |
10 to 60 cm H2O |
| PSV LImit |
10 to 60 cm H2O |
| Apnea Backup Time |
5 to 60 secs |
| Other Features |
Inspiratory Pause Limit, Sigh Breath, Manual Breath |
| |
| Alarms/Indications |
- Excess Breaths
- Low Expired Volume
- Low Inspired Volume
- FiO2 Sensor Open
- FiO2 Sensor Defective
- FiO2 High
- FiO2 Low
- High Inspiratory Pressure
- Low Inspiratory Pressure
- Patient Disconnection
|
- Ventilator Malfunction
- Low Inlet Volume
- Check Exp Flow Sensor
- Apnea
- Power Failure
- Peep Low
- Peep High
- Battery Operation
- Battery Low/ Battery Discharged
- Ventilator Failure
|
| |
|
| Monitored Parameters |
|
- FiO2 Monitoring
- Spontaneous Breaths
- Inspired Tidal Volume
- Expired Tidal Volume
- Peak Inspiratory Pressure
|
- Plateau Pressure
- PEEP Level
- Trigger Level
- Delivered BPM
|
| |
|
| Connections |
|
- 22 mm (Male) outlet port for tubing connection to patient.
- 22 mm (Male) expiratory port for tubing connection from patient.
- 8 mm low pressure connection for oxygen enrichment.
- Pressure monitoring port.
- Reservoir Bag Port
|
| |